Special Features
2hannel ratioing capability; high resolution display, 1 part in 3000; measurement speeds up to 40 pulses/second; porm 5% accuracy; statistical analysis of repetitive pulses: 10 or 100 pulse average, min & max energy and standard deviation; 64 character alphanumeric lcd dispay; digital compemsation for: window.Filter/beam splitter attenuator loss or normalizing detector response (range 0.1 to 1.0); interchangeanle window/filters/attenuators; auxiliary window/filters/attenuators; auciliary analog outputs
Width
10.000 inches
Length
12.000 inches
Ac Voltage Rating
115.0 volts and 240.0 volts
Frequency Rating
Between 50.0 hertz and 60.0 hertz
Electrical Power Source Relationship
Operating
Internal Battery Accommodation
Not included
Test Type For Which Designed
Optical energy measurements in joules
Operating Test Capability
14 decades 10 to the -12th (picajoules) to 100 joules depending on which of 8 probes are used
Phase
Single
Functional Description
Measures true energy/pulse from a single-shot to 40hz, picco joules to joules, uv to far-ir. Sets of 10 to 100 pulses can be statistically and analyzed, and the data can be transferred to a computer via an ieee48 interface bus. Accepts 2 probes and simutaneously displays the energy incident on channels a and b, as well the ratio of channel b/channel a. Maufactures name energy meter/ratiometer